Thermocouple Calibrator / Simulator

Thermocouple Calibrator / Simulator

The portable thermocouple calibrator is a compact and easy to use instrument that is equally at home in the laboratory or service environment. The DP6 will both measure and simulate the outputs from the 10 most popular thermocouple types, and may be scaled in °C, çF, K or mV. Automatic reference junction compensation is provided with the option for manual value entry.

For full portability the calibrator is powered by a rechargeable battery giving more than 15 hours continuous use. The DP6 is delivered in an ABS transport case which also houses the charger and a copper transition adapter for easy connection to miniature thermocouple plugs. A soft carrying bag with neck strap is available as an option.

Customer Comments

No comments have been made about this product. Be the first to let us know what you think.

Knowledge Base

You must be registered to view Knowledge Base articles related to this product. Please register or log-in to view these articles.

Key Features

  • Linearised for 10 Thermocouple Types
  • Measures and Sources
  • Calibrated to ITS90 and IPTS68
  • Displays in °C, °F, K and mV
  • 1mV Resolution
  • Reference Junction Compensation
  • Rechargeable Battery
  • Memory with 130 Stores
  • Digital Calibration

Downloads

Please feel free to download any of the documents associated with this product.

Technical Specification

 

THERMOCOUPLE TYPE

RANGE

ACCURACY

Code

Material

Degrees C

Measure & Source

B

PtRh30-PtRh6

+500 to +1820

± 0.5 °C

B

PtRh30-PtRh6

+200 to +500

± 1.5 °C

B

PtRh30-PtRh6

+60 to +200

± 6.0 °C

E

NiCr-CuNi

-200 to +1000

± 0.2 °C

E

NiCr-CuNi

-250 to -200

± 0.6 °C

E

NiCr-CuNi

-270 to -250

± 6.0 °C

J

Fe-CuNi

+800 to +1200

± 0.3 °C

J

Fe-CuNi

+200 to +800

± 0.2 °C

J

Fe-CuNi

0 to +200

± 0.1 °C

J

Fe-CuNi

-210 to 0

± 0.3 °C

K

NiCr-NiAl

+1000 to +1370

± 0.4 °C

K

NiCr-NiAl

+100 to +1000

± 0.3 °C

K

NiCr-NiAl

-50 to +100

± 0.1 °C

K

NiCr-NiAl

-150 to -50

± 0.2 °C

K

NiCr-NiAl

-225 to -150

± 0.5 °C

K

NiCr-NiAl

-270 to -225

± 3.0 °C

L

Fe-CuNi

+300 to +900

± 0.2 °C

L

Fe-CuNi

-100 to +300

± 0.1 °C

L

Fe-CuNi

-200 to -100

± 0.15 °C

N

NiCrSi-NiSi

+1100 to +1300

± 0.4 °C

N

NiCrSi-NiSi

+400 to +1100

± 0.3 °C

N

NiCrSi-NiSi

+150 to +400

± 0.15 °C

N

NiCrSi-NiSi

0 to +150

± 0.1 °C

R

PtRh13-Pt

+1200 to +1760

± 0.8 °C

R

PtRh13-Pt

+100 to +1200

± 0.4 °C

R

PtRh13-Pt

0 to +100

± 0.5 °C

R

PtRh13-Pt

-50 to 0

± 0.8 °C

S

PtRh10-Pt

+1400 to +1760

± 0.95 °C

S

PtRh10-Pt

+1200 to +1400

± 0.5 °C

S

PtRh10-Pt

+50 to +1200

± 0.4 °C

S

PtRh10-Pt

-50 to +50

± 0.6 °C

T

Cu-CuNi

-100 to +400

± 0.2 °C

T

Cu-CuNi

-230 to -100

± 0.5 °C

T

Cu-CuNi

-250 to -230

± 1.0 °C

T

Cu-CuNi

-270 to -250

± 2.5 °C

U

Cu-CuNi

+300 to +400

± 0.2 °C

U

Cu-CuNi

0 to +300

± 0.1 °C

U

Cu-CuNi

-150 to 0

± 0.15 °C

U

Cu-CuNi

-200 to -150

± 0.2 °C

Resolution on all types of thermocouple -0.1 deg.C, 0.1 K, 0.1 deg.F. Limits of error apply for 1 year at 20 degrees centigrade ±1 °C

RANGE

MAX. DISPLAY

UNCERTAINTY

RESOLUTION

10 mV

±15.000mV

±0.02% of reading ± 0.015% FS

1 mV

100 mV

±150.00 mV

±0.01% of reading ± 0.015% FS

10 mV

1 V

±1.5000V

±0.01% of reading ± 0.015% FS

100 mV

Temperature coefficient: typically 17 ppm/deg.C +0.2 mV /deg.C

Display 4 1/2 digit high contrast LCD 10.2 mm. Display range 19999 digits, automatic decimal point, polarity and units. Two line alphanumeric LCD for programming and display of configuration.
Terminals Two 4mm low thermal emf copper terminals.
Working Temperature 0... 40°C.
Storage Temperature -20... +50°C.
Battery  6 Volt 1.2 Ah sealed lead acid, replaceable.
Operating Time  15 Hours typical continuous use.
Charger Type  External charger operating from mains supply.
Case  Shockproof thermoplastic with polycarbonate sealed membrane keyboard.
Soft carrying case available as an accessory.

Reference Junction

 

Referenced to 0 Deg. C and with three operating modes. Automatic with internal sensor, Off (=to 0 °C), and Manual entry via keyboard
Accuracy  Better than ±0.1 °C at +20 °C.
Deviation  0.01°C / °C over the range 0 to +50 °C
Manual Input Range  The reference junction
reference value may also be set via the keyboard over the range 0 to +100
Size  50 x 130 x 60 mm (W D H)
350 x 260 x 65 mm in transport case.
Weight  1.4 kg

Knowledge Base

Product Matrix

The Product Matrix allows you to compare the features of up to 5 products. Simply check the box next to the products you wish to compare then select the 'Compare Products' button.

Product Enquiry

Contact Information

indicates required field
Title*
Please type in the validation code from the image (case sensitive).

Additional Information (optional)

 
Site engineered by NorthBloc